Commissioning
Fluid Components Intl
Document 06EN003327 Rev. -
51 FlexCOR
Model CMF Series
7.1 Keypad and display
layout
Keypad The keypad is used to set the flowmeter. The function of the keys are as follows:
TOP UP KEY This key (hold 2 sec.) is used to switch between operator
menu and setup menu. In the converter setup menu, a short
press will cause a return to the previous menu.
FORWARD KEY This key is used to step forward through the menus. It is the
only key normally used by the operator.
BACKWARD KEY This key is used to step backward through the menus.
CHANGE KEY This key changes the settings or numerical values.
SELECT KEY This key selects the figures to be changed.
LOCK/UNLOCK KEY This key allows the operator to change settings and gives
access to submenus.
Display
The display is alphanumerical and indicates flow values, flowmeter settings and error messsages.
The upper line is for primary flow readings and will always show either mass flowrate, volume
flowrate, density, temperature, totalizer 1 or totalizer 2. The line is divided into 3 fields.
S: Sign field
P: Primary field for numerical value
U: Unit field
The center line is the title line (T) with individual information according to the selected operator or
setup menu.
The lowest line is the subtitle line (ST) which either will add information to the title line or keep
individual information independent of the title line.
F: The alarm field. Two flashing triangles will appear by a fault condition.
M: The mode field. The symbols indicate the following.
Communication mode Basic settings
Service mode Output
Operator menu External input
Product identity Sensor characteristic
Language mode Reset mode
Ready for change Access to submenu (Press )
Value locked RESET MODE: Zero setting of
totalizers and initialization of setting
L: The lock field. Indicates the function of the lock key.
7. Commissioning
Zobrazit stránku 0
1 2 3 4 5 6 ... 24 25

Shrnutí obsahu

Strany 1 - ™ Model CMF Series

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 51 FlexCOR™ Model CMF Series7.1 Keypad and displaylayoutKeypad The keypad is us

Strany 2 - ™Model CMF Series 52

CommissioningFlexCOR™Model CMF Series 60Fluid Components IntlDocument 06EN003327 Rev. -7.5.3 External input7.5.2 Relay outputThe output functions err

Strany 3

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 61 FlexCor™ Model CMF SeriesMenu detail7.5.4 Sensorcharacteristi

Strany 4 - ™Model CMF Series 54

CommissioningFlexCor™ Model CMF Series 62Fluid Components IntlDocument 06EN003327 Rev. -7.5.5 Reset modeReset modeIn the reset mode menu the

Strany 5

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 63 FlexCor™ Model CMF SeriesExample; setting the0-pointKeypad op

Strany 6 - ™Model CMF Series 56

CommissioningFlexCOR™Model CMF Series 64Fluid Components IntlDocument 06EN003327 Rev. -7.5.6 Service modeAll previous settings are re-initialized whe

Strany 7

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 65 FlexCOR™ Model CMF SeriesService modeThe service mode menu can be used to ch

Strany 8 - ™Model CMF Series 58

CommissioningFlexCOR™Model CMF Series 66Fluid Components IntlDocument 06EN003327 Rev. -Menu setup7.5.7 Operator menusetupThe upper line is always act

Strany 9

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 67 FlexCOR™ Model CMF Series7.5.8 Product identitySoftware version of add-on mo

Strany 10 - ™Model CMF Series 60

CommissioningFlexCOR™Model CMF Series 68Fluid Components IntlDocument 06EN003327 Rev. -7.5.11 HART®communication(Add-on module)Menu setup

Strany 11

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 69 FlexCOR™ Model CMF Series7.6.3 BatchA batch can be started, paused or stoppe

Strany 12 - Commissioning

CommissioningFlexCOR™Model CMF Series 52Fluid Components IntlDocument 06EN003327 Rev. -7.2 Menu build-upThe menu structure of a specific type of tran

Strany 13

CommissioningFlexCOR™Model CMF Series 70Fluid Components IntlDocument 06EN003327 Rev. -Parameter Factory setting Settings availablePasswordPassword 1

Strany 14 - ™Model CMF Series 64

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 71 FlexCOR™ Model CMF SeriesDirection/limit Off Mass flow, fraction A, fraction

Strany 15

CommissioningFlexCOR™Model CMF Series 72Fluid Components IntlDocument 06EN003327 Rev. -Error systemThe converter system is equipped with an error and

Strany 16 - ™Model CMF Series 66

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 73 FlexCOR™ Model CMF Series7.8.2 List of errornumbersError Error text #Comment

Strany 17

T. s.FlexCOR™ Model CMF Series 74Fluid Components IntlDocument 06EN003327 Rev. -8.1 TransmitterSymptom Output Error Cause Remed

Strany 18 - ™Model CMF Series 68

T. s.Fluid Components IntlDocument 06EN003327 Rev. - 75 FlexCOR™ Model CMF Series8.2 Check for air in thesystemIn case of la

Strany 19

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 53 FlexCOR™ Model CMF Series

Strany 20 - ™Model CMF Series 70

CommissioningFlexCOR™Model CMF Series 54Fluid Components IntlDocument 06EN003327 Rev. -7.4 Basic settingsMenu detail

Strany 21

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 55 FlexCOR™ Model CMF SeriesBasic settingsMenu descriptionSetting of min./max.v

Strany 22 - ™Model CMF Series 72

CommissioningFlexCOR™Model CMF Series 56Fluid Components IntlDocument 06EN003327 Rev. -Setting the low-flow cut-offIn certain applications flow sign

Strany 23

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 57 FlexCOR™ Model CMF SeriesMenu detailDigital outputFrequency7.5.1 Outputs set

Strany 24 - 8. Trouble shooting

CommissioningFlexCOR™Model CMF Series 58Fluid Components IntlDocument 06EN003327 Rev. -Menu detailDigital outputLimit/directionLimit switches are ava

Strany 25

CommissioningFluid Components IntlDocument 06EN003327 Rev. - 59 FlexCOR™ Model CMF SeriesDigital outputBatch menuMenu detailBatch menuBatchi